This website uses cookies

Read our Privacy policy and Terms of use for more information.

Most coating inspectors have been handed a gage at some point and told, "It's calibrated."

That may be true. It doesn't mean the instrument is ready for the exact surface, coating system, or field condition sitting in front of you that day.

Every reading you write down is only as good as the instrument that produced it. A thickness gage that has drifted, a profile gage that was never zeroed, or a holiday detector set to the wrong voltage can turn a passing job into a failure, or pass a job that should have failed. Bad information is worse than no information, because it gives everyone confidence in numbers that may not be right.

Certified Calibration, Field Verification, and Field Adjustment

Certified calibration is the traceable calibration performed by the manufacturer or a qualified calibration lab. That is the certificate and sticker showing the instrument was checked against known standards under controlled conditions.

Field verification is what the inspector does before relying on the instrument. The inspector checks the gage against a known standard to confirm it is reading properly in the field.

Field adjustment is when the inspector adjusts the instrument, if the manufacturer allows it, so it reads correctly on the actual substrate, thickness range, or test condition being inspected.

All three matter, and none of them replaces the others. Skipping any one can turn good equipment into bad data.

Certified Calibration Is the Starting Point

Most coating inspection instruments should carry a current certificate of calibration from the manufacturer or a qualified calibration lab. On many projects, that means annual certified calibration.

The certificate matters because it provides traceability. It shows the instrument was checked against known standards under controlled conditions, and it gives the inspector, contractor, owner, and auditor something objective to rely on if the readings are ever questioned.

A DFT gage, surface profile gage, durometer, or holiday detector can hold a valid certificate and still give poor results if it is not verified, adjusted, grounded, cleaned, or used correctly.

The certificate answers one question: was the instrument calibrated? The field check answers a different one: is it giving reliable readings here, on this work? Both questions matter.

"A current calibration sticker doesn't make the inspector right. It makes the instrument traceable."

DeFelsko PosiTector 6000: Dry Film Thickness

The 6000 is one of the most common thickness gages in the field, and one of the easiest tools to trust too quickly.

The right way to set it up is to zero the gage on clean, blasted steel representative of the surface being inspected, so the zero accounts for the surface profile. Then place certified shims over that same blasted surface to field verify. This confirms the gage reads correctly over the real substrate, not over a smooth plate that may behave nothing like blasted steel.

Bracket the expected thickness when you verify. If the specification is 16 to 20 mils, do not verify with only a 2 mil shim. Use shims near your working range. The gage should read within the manufacturer's stated tolerance for that probe and range. If it is out, use the 1-point or 2-point cal adjustment against the shims, if allowed by the manufacturer and your procedure.

Metallurgy, curvature, and blast profile can all affect the reading, which is why zeroing and verifying on representative blasted steel matters.

A practical field routine:

  • Check the calibration due date.

  • Inspect the probe face and cable, and confirm the correct probe for the substrate.

  • Zero the gage on clean, blasted steel representative of the work being inspected.

  • Verify with certified shims placed over that blasted steel, near the expected coating thickness.

  • Bracket the expected range with low and high shims when possible.

  • Record the verification results.

  • Recheck during the shift, after a drop, a battery change, a large temperature swing, or anytime the readings stop making sense.

Field verification is not just a morning paperwork step. It is part of controlling the inspection process.

Surface Profile Gage: PosiTector SPG

Surface profile is not coating thickness, and the gage should not be checked like a coating thickness gage.

The SPG measures peak-to-valley surface profile with a spring-loaded pin that drops into the valleys while a flat base rests on the peaks. This is the depth micrometer method described in ASTM D4417 Method B. It is a different measurement than dry film thickness, and the most common mistake is treating it like a coating thickness gage.

Coating thickness shims are not the correct standard for verifying an SPG. A depth micrometer profile gage has to be checked against standards made for surface profile verification, following the manufacturer's instructions.

Before use, confirm the pin is clean and undamaged and the base sits flat. Zero the instrument on the smooth glass zero check plate, then verify against the proper certified profile standard. Take several readings across the blasted surface and average them per your specification, because profile varies from spot to spot.

Profile matters because it affects adhesion and coating coverage over the peaks. A bad profile reading can accept a surface that is too smooth, reject an acceptable one, or create confusion when coverage becomes an issue later. Bad profile readings lead to bad coating decisions.

Durometer: Shore D Hardness

We use the Shore D durometer to check hardness on hard cured coatings and linings where hardness is part of the acceptance criteria. ASTM D2240 is commonly referenced for this type of testing. Confirm up front that Shore D is what the project specification or coating manufacturer calls for, since that is what we verify and report against.

  • Inspect the indenter tip and presser foot for wear, flat spots, or damage.

  • Verify the gage against the supplied Shore D test block or a certified reference block before field use.

  • Apply firm, steady pressure and read at the specified dwell time.

  • Take several readings and average them as required by the specification or procedure.

Remember what the number depends on. Shore D readings can be affected by temperature, coating thickness, substrate support, surface texture, and cure condition. A reading on a thin coating over hard steel is not the same as one taken on a thick lining.

The biggest mistake is using the number without context. Document the instrument, Shore D reference block, location, material condition, temperature when relevant, dwell time if required, and the acceptance criteria being used.

Holiday Detectors: Voltage Verification Is Not Optional

Holiday detection is one of the most important steps on coated pipe, tanks, and linings, and one of the most misunderstood.

A high voltage detector is not simply turned on and used. The two things that ruin a survey are the wrong voltage and a bad ground, so check both first.

Set the voltage from the coating's measured thickness. Some inspectors still use older rules of thumb, such as roughly 100 to 125 volts per mil, but those shortcuts should never override the approved procedure. AMPP/NACE SP0188 is frequently specified, and when it is, calculate the voltage from its formulas and tables for the measured thickness. ASTM D5162, ASTM G62, manufacturer instructions, and project-specific procedures may also apply depending on the coating system and work being performed.

Too much voltage can burn through sound coating and create the very defect you are looking for. Too little voltage can miss defects. Follow the approved procedure and the manufacturer's data.

How to Field Verify a Holiday Detector

Field verifying a holiday detector is simpler than it may sound, but it is an important step before production testing begins.

Start by setting the holiday detector voltage for the coating system and thickness being tested. In many field situations, the voltage is based on the average coating thickness. For example, if the coating is running between 30 and 40 mils, the voltage would typically be set using 35 mils as the basis, unless the project specification or approved procedure requires something different.

Once the voltage is set, verify the actual output at the probe. One common tool for this is the SPY Model PJM, often referred to as a pocket jeep meter. Connect the PJM's black cable to the holiday detector ground and the red cable to the probe. Turn the verifier on, energize the holiday detector, and read the actual voltage output at the probe.

The verified voltage should match the voltage setting within the allowed tolerance, commonly about 10 percent unless the specification or manufacturer states otherwise. If the reading is outside tolerance, adjust the holiday detector and check it again. Do not adjust the verifier to match the detector.

Some newer detectors, including the DeFelsko HHD, have a built-in voltmeter or verification feature that helps confirm output voltage. Whether the voltage is checked with a separate verifier or through the detector's built-in feature, the goal is the same: confirm the voltage at the electrode before testing the coating.

As a final check, pass the electrode over a known holiday on a test piece or coupon, when allowed by the procedure. The detector should spark and alarm at the defect. That confirms the voltage, ground path, electrode contact, and alarm function are working before the inspection begins.

Safety matters too. High voltage detectors carry a real shock hazard. Follow the manufacturer's safety instructions, keep bystanders clear, and never test around flammable vapors.

Field Verification Should Be Documented

If it is not documented, it is hard to defend later.

Inspection reports should include the instrument make, model, serial number, and calibration due date, along with the verification standard used, the verification readings, the required tolerance, and the time of the check.

For holiday detectors, also record the voltage setting, the coating thickness basis, the electrode type, the grounding method, and the applicable standard or procedure.

It does not need to be complicated, but it does need to be consistent. A simple daily verification log shows control of the inspection process. It shows that the inspector did not just show up with a gage and start writing numbers down. It also protects the inspector when the data is questioned months later.

Why Yearly Certified Calibration Still Matters

Some people treat annual calibration as a paperwork requirement. It does more than that.

Certified calibration catches drift, damage, wear, probe issues, and electronic problems that are not obvious in daily use, and it provides traceability back to recognized standards. That traceability gives your data weight when a project is audited, disputed, or reviewed after a failure.

A field shim check is not certified calibration. A jeepmeter check is not certified calibration. A test block check is not certified calibration. Those field checks are necessary, and they work together with the controlled calibration process, but they do not replace it.

The certified shims, blocks, and verifiers you check against also have to be recertified on schedule, because a worn standard makes every field check meaningless.

The Real Goal

The goal is not to pass a paperwork audit. It is to produce inspection data that can be trusted.

When a DFT gage reads 18 mils, the owner should be able to trust that number. When an SPG reads 3.5 mils, the applicator should be able to trust it. When a Shore D reading is used to evaluate cure, everyone should know the method and test conditions were right. When a holiday detector passes over coated pipe without alarming, the project should have confidence that the voltage, grounding, electrode contact, and function were all correct.

Proper calibration, field verification, adjustment, and documentation are how the inspector proves the equipment was ready to do its job. The numbers matter, and the process behind the numbers matters just as much.

"Inspection equipment doesn't make good decisions on its own. The inspector does."

How does your crew handle daily verification before production testing starts? We'd like to hear what's working in the field.

Summary & Key Takeaways

  • Three checks, three jobs. Certified calibration proves traceability, field verification proves the reading is good here, and field adjustment sets the gage to the real substrate and range. None of them replaces the others.

  • A certificate is not a field check. A current calibration sticker makes the instrument traceable. It does not prove the gage reads right on this surface, this coating, this day.

  • DFT: zero on blasted steel, verify with shims over it. Zero the PosiTector 6000 on clean blasted steel representative of the work, then bracket the specified range with certified shims placed over that steel, not over a smooth plate.

  • Surface profile is not coating thickness. Verify the SPG against a certified profile standard by the depth micrometer method (ASTM D4417 Method B). Coating thickness shims are the wrong standard.

  • Holiday detectors: set the voltage, then prove it. Calculate voltage from the measured thickness using SP0188, confirm the actual output at the probe with a pocket jeep meter or a built-in verifier, and check the ground. The wrong voltage or a bad ground ruins the survey.

  • If it is not documented, it did not happen. Record instrument make, model, serial, and cal due date, plus the standard used, the readings, the tolerance, and the time. A simple daily log shows control of the process and defends the data later.

Referenced Standards & Technical Resources

  • AMPP/NACE SP0188-2024 — Discontinuity (Holiday) Testing of New Protective Coatings on Conductive Substrates. Voltage selection and holiday testing procedure.

  • ASTM D4417 — Standard Test Methods for Field Measurement of Surface Profile of Blast Cleaned Steel; Method B (depth micrometer) is the basis for the SPG.

  • ASTM D2240 — Standard Test Method for Rubber Property, Durometer Hardness (Shore D).

  • ASTM D5162 — Discontinuity (Holiday) Testing of Nonconductive Protective Coating on Metallic Substrates.

  • ASTM G62 — Holiday Detection in Pipeline Coatings.

  • AMPP/NACE Coating Inspector Program (CIP) Level 1 — Surface preparation instrumentation, dry film thickness instrumentation, and high- and low-voltage holiday testing chapters.

  • Manufacturer instructions — DeFelsko PosiTector 6000 and SPG operating manuals for probe selection, zeroing, cal adjustment, and verification tolerances.

Roberts Corrosion Services, LLC

Established in 2011, Roberts Corrosion Services, LLC delivers comprehensive, turn-key cathodic protection and corrosion control solutions nationwide. Our end-to-end expertise encompasses design and inspection, installation and repair, surveys and remedial work. We provide drilling services for deep anode installations and a full laboratory for analysis of samples and corrosion coupons, as well as custom CP Rectifier manufacturing.

While our initial focus was on the Appalachian Basin area, we complete field work all over the US. We are a licensed contractor in many states and can complete a wide range of services.

Our biggest strength is in our flexibility for our clients. Solutions and Results.

Let us know how we can help.

Location: 39.251882, -81.047440

(304) 869-4007

Keep Reading